Safety considerations
Invenia ABUS 2.0 – System Setup and Basic Service Manual 1-17
4700-0043-00 Rev. 4
Electrical safety
Safe practices
Follow these guidelines to minimize shock hazards whenever
you are using the Invenia ABUS 2.0:
The Invenia ABUS 2.0 is equipped with a three-conductor
AC power cable. This must be plugged into an approved
electrical outlet with third wire safety ground.
The power outlet used for this equipment should not be
shared with other types of equipment.
Both the system power cable and the power connector must
meet international electrical standards
Transducer
Follow these guidelines before connecting the transducer to the
Invenia ABUS 2.0:
Inspect the transducer prior to each use for damage or
degradation to the:
• housing
• lens
• seal
• connector pins
Do not use a damaged or defective transducer.
Never immerse the transducer connector or adapter into
any liquid.
DANGER
Electrical Warning
Connecting an Invenia ABUS 2.0 to the wrong voltage level will
most likely damage it.