EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ E10

GE LOGIQ E10 Service Manual

GE LOGIQ E10
778 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #146 background imageLoading...
Page #146 background image
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
4 - 16 Section 4-2 - Functional checks
Recommended Test Phantoms
GE recommends the RMI 430GS phantom is preferred, but it is not required. It is the most current
phantom recommended to our field service personnel and provides the necessary targets and extended
life necessary for consistent Ultrasound system testing.
Figure 4-7 Performance tests
2
4321
4
6
8
10
12
14
16
RMI 403GS
Lateral Distance
Measurement
Functiona
l
Resolutio
n
Contrast
Resolution
&
Gray Scale
P
hotography
Lateral
Resolution
Axial
Resolution
Axial Distance
Measurement
Penetration

Table of Contents

Other manuals for GE LOGIQ E10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E10 and is the answer not in the manual?

GE LOGIQ E10 Specifications

General IconGeneral
BrandGE
ModelLOGIQ E10
CategoryMedical Equipment
LanguageEnglish

Related product manuals