DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 8 Replacement Procedures 8 - 33
8-6-3 Side Cover(s)/Side Air Intake (Filter) inspection / replacement
NOTE: When replacing Side Covers for a LOGIQ E10 used in veterinary environment, the LOGIQ E10
Spare Parts Handling for Veterinary Systems MUST BE re-installed before the LOGIQ E10
returned for use. See: "Label Placement for the LOGIQ E10 used in a Veterinary Environment"
on page 8-58.
The Side Covers must be removed before any lower console cover(s) can be removed.
Table 8-21 Manpower / Time and Tools
Manpower /
Time
Tools / PPE
One person /
5 minutes per
Side Cover
Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
8-2-6 "PPE Required During Service" on page 8-6.
Table 8-22 Preparations and Preparation Link
Preparations - you must perform the following steps
1. Power down the system.
2. Disconnect the mains power cable from the wall outlet.
Preparation Link (if you need more information):