DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 4-7 - Keyboard Function Checks 4-31
8.) Check the 'Fast' radio button into Indicator box on scfpapitest program and press 'Set All' button.
R (Right)
Right Button
BodyPattern/Ellipse (Rotary
BodyPattern/Ellipse PushEncoder
Rotary
THI
Harmonics Button
Comment comment Button
AO
Auto Right Button
Clear Clear Button
B (Button)
B PushEncoder Button
TGC Slide Slider : n=8 s1 ~ s8
B (Rotary)
B PushEncoder Rotary
NOTICE
When you can check the name displayed in log as you press or rotate keys, the main keyboard assy
was replaced correctly.
Figure 4-30 Fast button
NOTICE
After pressing Fast radio button, indicators should be blanked on the frontpanel.
Table 4-8 Button and Rotaty Table
Button or Rotary Name The Name displayed in Log Button or Rotary Name The Name displayed in Log