EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ P5 Basic OB

GE LOGIQ P5 Basic OB User Manual

GE LOGIQ P5 Basic OB
532 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #248 background imageLoading...
Page #248 background image
GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
8-2 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
Section 8-2 DISASSEMBLY/RE-ASSEMBLY
NOTE: If the LOGIQ P5/A5/A5pro needs to be sent for repair, ensure that any patient information is
erased from the storage device, or that the storage device is removed from the LOGIQ P5/A5/
A5pro before shipping. In case that any patient information is still residing on the LOGIQ P5/
A5/A5pro, GE will contact the customer and request for urgent collection of that patient
information. GE will keep this patient information in a secure environment for a maximum period
of 1 month. All patient information will be permanently deleted at that point.
If PHI (Patient Healthcare Information) data needs to be sent to GE employees for service purposes,
GE will ascertain agreement from the customer. The patient information shall only be transferred by
approved service processes, tools and devices restricting access, protecting or encrypting data where
required, and providing traceability in the form of paper or electronic documents at each stage of the
procedure while maintaining compliance with cross-border restrictions of patient information
WARNINGWARNING
ONLY QUALIFIED SERVICE PERSONNEL SHOULD REMOVE ANY COVERS OR
PANELS. ELECTRICAL HAZARDS EXISTS AT SEVERAL POINTS INSIDE. BECOME
THOROUGHLY FAMILIAR WITH ALL HAZARDOUS VOLTAGES AND HIGH CURRENT
LEVELS TO AVOID ACCIDENTAL CONTACT
CAUTION
Do not wear the ESD wrist strap when you remove a part of power supply unit. Turn OFF power
and unplug the power cord before removing a part of power supply unit. However be sure to turn
off power and wear the strap before you remove a circuit boards.
WARNINGWARNING
DO NOT SERVICE OR DISASSEMBLE PARTS UNDER FRU UNIT LEVEL AT ANY
CIRCUMSTANCES.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 Basic OB and is the answer not in the manual?

GE LOGIQ P5 Basic OB Specifications

General IconGeneral
BrandGE
ModelLOGIQ P5 Basic OB
CategoryMedical Equipment
LanguageEnglish

Related product manuals