EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ P5 Basic OB

GE LOGIQ P5 Basic OB User Manual

GE LOGIQ P5 Basic OB
532 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #215 background imageLoading...
Page #215 background image
GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 7-3 - Screen Captures 7-7
7-3-3 Capturing a Screen (cont’d)
5.) A Save dialog box will be opened. Choose the destination as the archive location to save the
image on the CD-R/DVD-R, USB or network storage.
7-3-4 Reset the P2 Key to Customer’s Functionality
If the customer had programmed the P2 Key to a function other than screen capture, restore that
functionality recorded in section 7-3-1 on page 4. refer to Figure 7-2.
1.) Select
Utility from the Keyboard.
2.) Select
Connectivity from the Menu.
3.) Select the
Buttons tab on the Connectivity screen.
4.) In the
Button field, select Print2.
5.) In the “Physical Print Buttons”box, select the options recorded in step 5, Section 7-3-1. :
* Format
* Image Frames
* Capture Area
* Compression
Figure 7-5 Save Dialog Box