GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
5-34 Section 5-7 - Power Diagrams
5-7-4 AC Power Outlet for Peripheral
The ARP CONN ASSY and ARP CONN 220V ASSY are the ARP assemblies which interconnect
external signal and power with the LOGIQ™ P5 system. Both of the assemblies are the same except
the circuit breaker's current ratings depending on voltage of the system power source (See 'Safety'
described below).
5-7-4-1 Safety
Circuit Breaker
Electrical Isolation
The ARP's Insite interface is electrically isolated from the system's internal circuit. The isolator's
dielectric strength should be 500VAC RMS or higher within 1 minute (in compliance with IEC60601-1-
1).
Table 5-4 Circuit Breaker Specification
Assembly
Power Voltage
Rating (V)
Peripheral Current
Rating (A)
Trip Time Reset Time
ARP CONN
110
4.0
1 hour max. @
5.4A, 25 degreeC
110 second
ARP CONN 220V
220
2.0
1hour max. @
3.5A, 25 degreeC
180 second

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 Basic OB and is the answer not in the manual?

GE LOGIQ P5 Basic OB Specifications

General IconGeneral
BrandGE
ModelLOGIQ P5 Basic OB
CategoryMedical Equipment
LanguageEnglish

Related product manuals