DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-89
8-2-36 Battery on SYSCONPM/CM Assy
NOTE: If the battery on SYSCONPM/CM Assy runs out, the system date will be reset. Then can not
exit to maintenance mode.
NOTE: The Battery is UL critical component. So must use the battery approved. The recommended
one is CR2450 manufactured by TOSHIBA.
8-2-36-1 Tools
Common pilIips screwdrivers
Option strings
8-2-36-2 Preparations
Shut down the system and switch off the main breaker.
8-2-36-3 Removal procedure
1.) Remove the Side Right Cover. Refer to the 8-2-16 "Side Right Cover" on page 8-52.
2.) Remove the EMI Cover R. Refer to the 8-2-27 "EMI Cover R" on page 8-69.
3.) Unplug the J13 connector. Refer to the Figure 8-123.
4.) Remove the SYSCONPM(SYSCONCM) assy in slot #3 in the nest box.
Figure 8-123 Pull out the SYSCONPM(SYSCONCM) Assy
Unplug the J13 Connector