GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
8-90 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-36-3 Removal procedure (cont’d)
5.) Remove the battery.
6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-36-4 Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-124 Pull out the battery from SYSCON PWA
Table 8-37 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
5394141, Rev 1+, Section 8-2-36. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-9-4
SYSCONML Assy Function Validation
Procedure
Section 4-3-4
System B/M-Mode Checks
Section 4-3-5
System CFM and PWD Checks
Section 4-5
Peripheral Checks
Section 10-5-2
Functional Checks (See Also Chapter 4)
Section 10-5-2-2
Peripheral/Option Checks
Section 10-5-5
Physical Inspection

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 Basic OB and is the answer not in the manual?

GE LOGIQ P5 Basic OB Specifications

General IconGeneral
BrandGE
ModelLOGIQ P5 Basic OB
CategoryMedical Equipment
LanguageEnglish

Related product manuals