DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
8-60 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-20 Top Cover
8-2-20-1 Tools
Common pilIips screwdrivers
8-2-20-2 Preparations
Shut down the system and switch off the main breaker.
8-2-20-3 Removal procedure
1.) Remove the Dummy Cover L/R. Refer to the 8-2-25 "Dummy Cover L / R" on page 8-66.
2.) Remove the Top Bottom Cover. Refer to the 8-2-21 "Top Bottom Cover" on page 8-60.
3.) Remove 2 rubber caps (1-2) from the top cover of the system and unscrew 2 screws (1-2).
4.) Unscrew 5 screws (1-5) from the bottom of the top cover.
5.) Remove the Top Cover.
6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-20-4 Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-95 2 rubber caps of top cover
Table 8-21 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
5394141, Rev 1+, Section 8-2-20. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-8-1
Cover Parts Function Validation
Section 10-5-5
Physical Inspection
(1)
(2)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 Basic OB and is the answer not in the manual?

GE LOGIQ P5 Basic OB Specifications

General IconGeneral
BrandGE
ModelLOGIQ P5 Basic OB
CategoryMedical Equipment
LanguageEnglish

Related product manuals