GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 10-6 - Electrical Safety Tests 10-13
10-6-2 GEMS Leakage Current Limits
The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment
Safety Standards. These limits are GEMS standards and in some cases are lower than the above
standards listed.
NOTE: *Mains Applied refers to the sink leakage test where mains (supply) voltage is applied to the
part to determine the amount of current that will pass (or sink) to ground if a patient contacted
mains voltage.
The following tests are performed at the factory and should be performed at the site. These tests are:
grounding continuity, chassis leakage current, probe leakage current, and ECG leakage current. All
measurements are made with an electrical safety analyzer Model 600/600E built by Dale Technology
Corporation or equivalent device.
Table 10-14 Chassis Leakage Current Limits—Accessible Metal Surfaces
Country Normal Condition Open Ground Reverse Polarity Open Neutral
USA
N/A
0.3 mA 0.3 mA N/A
Other
0.1 mA
0.5 mA 0.5 mA 0.5 mA
Table 10-15 Type BF Applied Part Leakage Current Limits - Non-Conductive (Floating) Surface
and Cavity Probes
Country
Normal
Condition Open Ground Reverse Polarity Open Neutral *Mains Applied
USA
0.05 mA
0.05 mA 0.05 mA 0.05 mA N/A
Other
0.1 mA
0.5 mA 0.5 mA 0.5 mA 5.0 mA
Table 10-16 Type CF Applied Part Leakage Current Limits - Surgical Probes and ECG
Connections
Country
Normal
Condition Open Ground Reverse Polarity Open Neutral *Mains Applied
USA
0.01 mA
0.05mA 0.05 mA N/A 0.025 mA
Other
0.01 mA
0.05 mA 0.05 mA 0.05 mA 0.05 mA

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 Basic OB and is the answer not in the manual?

GE LOGIQ P5 Basic OB Specifications

General IconGeneral
BrandGE
ModelLOGIQ P5 Basic OB
CategoryMedical Equipment
LanguageEnglish

Related product manuals