GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
8-56 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-18 Rear Cover
8-2-18-1 Tools
Common pilIips screwdrivers
8-2-18-2 Preparations
Shut down the system and switch off the main breaker.
8-2-18-3 Removal procedure
1.) Remove the Left, and Right side covers. Refer to the 8-2-15 "Side Left Cover" on page 8-51 and 8-
2-16 "Side Right Cover" on page 8-52.
2.) Remove 4 rubber caps from the rear cover of the system.
3.) Unscrew 4 screws (1-4) from the rear cover.
4.) Remove the Rear Cover.
Figure 8-91 4 rubber caps on Rear Cover
Figure 8-92 4 screws on rear cover
(2)
(3)
(4)
(1)