Replacement Procedures
7-16 LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
Loading the System Software with USB memory stick (For R1.0.x)
This section describes how to upgrade the software upgrade
between R1.0.x.
1. Insert the USB memory stick labeled “System & Application
Software” to the system.
2. Properly turn off the scanner by momentarily pressing the
Power On/Off Switch. In System-Exit window, select
Shutdown to shutdown the system.
NOTE: If the system will not shutdown normally, hold down the
Power On/Off Switch until the light turns off.
3. Power on the system. The system will detect the USB
memory stick automatically.
4. Select one of the options and then press Enter key to
continue. Select choice [1] to load the C disk only.
Figure 7-2. System Software load instruction