Loading the software
LOGIQ V2/LOGIQ V1 – Basic Service Manual 7-17
5610739-100 English Rev.10
Loading the System Software with USB memory stick (For R1.0.x)
(continued)
• To select [1], only the bootable C: partition is loaded.
This option is intended for recovery of a system that will
not boot up. It is not recommended for application
software upgrade because during upgrade process, the
data on the system would possibly be impacted.
• To select [2], wipe tool: This option is to wipe all the
patient data.
NOTE: All the patient data cannot be recovered after wipe
process.
• To select [3], Service tool. The complete disk will be
loaded. This option is recommended for application
software upgrade.
NOTE: This option is used by service engineer only. It needs
password to continue, the password is GEWuXiChina
and it is case sensitive.
NOTE: When to select [3] to load complete disk, please ensure
that any patient data on the disk has been backed up.
CAUTION
If you select [3], ALL existing software and data will be erased.
If backup has not been performed, all data like Patient
Database, System Configuration and User Configurations
(Customer Presets) will be lost.
WARNING
While the software install procedure is designed to
preserve data, you should select choice [1] to format disk
C only.

Table of Contents

Other manuals for GE LOGIQ V2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V2 and is the answer not in the manual?

GE LOGIQ V2 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V2
CategoryMedical Equipment
LanguageEnglish

Related product manuals