Revision B 3-3
Maintenance: Introduction
MAC 5000 resting ECG analysis system
2000657-002
Introduction
Recommended
Maintenance
Regular maintenance, irrespective of usage, is essential to ensure that
the equipment will always be functional when required.
Preventive Maintenance
Inspection Report
To help you establish a systematic maintenance routine, GE Marquette
Medical Systems recommends that, every six months, you perform the
maintenance checks and test procedures on the “Preventive
Maintenance Inspection Report,” included at the end of this chapter.
Required Tools and
Supplies
In addition to a standard set of hand tools, you will need the items listed
below.
NOTE
Unless you have an Equipment Maintenance
Contract, GE Marquette Medical Systems
does not in any manner assume the
responsibility for performing the
recommended maintenance procedures.
The sole responsibility rests with the
individual or institution using the equipment.
GE Marquette Medical Systems service
personnel may, at their discretion, follow the
procedures provided in this manual as a
guide during visits to the equipment site.
Table 3-1. Tools and Supplies
Item Part Number
#10 TORX driver
Leakage current tester MT-1216-02AAMI (for 220V)
MT-1216-01AAMI (for 110V)
Multifunction micro-simulator MARQ 1
Precision dust remover
Lint-free soft cloth TX609

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Marquette MAC 5000 and is the answer not in the manual?

GE Marquette MAC 5000 Specifications

General IconGeneral
BrandGE
ModelMarquette MAC 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals