PACE Series SCPI Manual
K0472 Revision A 4-87
:SOUR:PRES:INL:TIME
Applicability: PACE5000, PACE6000
Command Syntax
:SOURce[x][:PRESsure]:INLimits:TIME <number>
where x = 1 or 2 is the module number
(default - 1)
Parameter: <integer> in-limits time in seconds
Short form: :SOUR:INL:TIME <number>
Function: Sets the time that the pressure has to be within limits before
generating a service request.
Default: 2 seconds
minimum 2 seconds
maximum 999 seconds
Query Syntax
:SOURce[x][:PRESsure]:INLimits:TIME?
where x = 1 or 2 is the module number
(default - 1)
Short form: :SOUR:INL:TIME?
Function: Query in-limits timers.
Response: Number representing in-limits time in seconds.
Description
Sets the in-limits timer value.
Example
TX> :SOUR:PRES:INL:TIME?
RX> :SOUR:PRES:INL:TIME 2
Can be set to 99 by issuing command:
TX> :SOUR:PRES:INL:TIME 99
TX> :SOUR:PRES:INL:TIME?
RX> :SOUR:PRES:INL:TIME 99
Or to 999 by issuing command:
TX> :SOUR:PRES:INL:TIME 999
TX> :SOUR:PRES:INL:TIME?
RX> :SOUR:PRES:INL:TIME 999