PACE Series SCPI Manual
K0472 Revision A 4-33
:OUTP:LOG
Applicability: PACE1000, PACE5000, PACE6000
Note: The VFC option must be fitted.
Command Syntax
:OUTP[x]:LOGic[y]
where: x (module) = 1 or 2 (default - 1) and y = 1, 2 or 3 represents the relay number (default - 1).
Parameter: <boolean>
OFF - turn relay [y] OFF
ON - turn relay [y] ON
[y] = 1, 2 or 3
Short form: :OUTP:LOG
Function: Turns relay ON and OFF
Default: Relay 1 of module 1
Query Syntax
where: x (module) = 1 or 2 (default - 1) and y = 1, 2 or 3 represents the relay number (default - 1).
:OUTP[x]:LOGic[y]?
Short form: :OUTP:LOG?
Function: Asks for relay ON/OFF status.
Response: 0 - relay OFF
1 - relay ON
Description
With the volt-free contact option installed and each relay set to ‘communication’ through the
front panel control. The three relays can be queried and switched on and off. Omitting [x] or
[y], the default of 1 is used.
These commands return an error message: “The option board is not configured correctly”, if
the volt-free contact option is installed but the relay not set to ‘communication’.