PACE Series SCPI Manual
K0472 Revision A 4-55
:SENS:PRES:SLEW
Applicability: PACE1000, PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
:SENSe[x][:PRESsure]:SLEW?
where x = 1 or 2 is the module number
(default - 1)
Short form: :SENS:SLEW?
Function: Asks for current slew rate.
Response: Decimal number representing slew rate always in current pressure units
per second.
Description
This query gets the slew rate of the input pressure always in
current pressure units per
second.
A constant input pressure gives a slew rate of: 0.0.
Example
1
TX> :SENS:PRES:SLEW?
RX> :SENS:PRES:SLEW -20.1089802
Example
2
TX> :SENS2:PRES:SLEW?
RX> :SENS2:PRES:SLEW -20.0536633