PACE Series SCPI Manual
K0472 Revision A 4-35
:OUTP:ISOL:STAT
Applicability: PACE5000, PACE6000
Command Syntax
:OUTPut[x]:ISOLation:STATe
where x = 1 or 2 is the module number
(default - 1)
Parameter: <boolean>
0 OFF - turn isolation valve of the module [x] OFF
1 ON - turn isolation valve of the module [x] ON
Short form: :OUTP:ISOL:STAT
Function: Turns isolation valve ON and OFF
Default: -
Query Syntax
:OUTPut[x]:ISOLation:STATe?
where x = 1 or 2 is the module number
(default - 1)
Short form: :OUTP:ISOL:STAT?
Function: Asks for isolation valve ON/OFF status.
Response: 0 - isolation valve OFF
1 - isolation valve ON
Description
The isolation valve of the module [x] can be queried and switched on and off.