PACE Series SCPI Manual
K0472 Revision A 4-43
SENS:AIRF:QNH
Applicability: PACE1000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
SENSe:AIRField:QNH?
Short form: SENS:AIRF:QNH?
Function: This query reads the QNH value when the instrument is in airfield mode.
Response: A decimal reading in selected units of pressure.
Description
Queries the QNH reading, returns the QNH reading in selected units of pressure when the
instrument is in airfield mode.
Note: QNH is the pressure measured at a specific site then reduced to mean sea level
pressure.