PACE Series SCPI Manual
K0472 Revision A 4-49
:SENS:PRES
Applicability: PACE1000, PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
:SENSe[:PRESsure]?
where x = 1 or 2 is the module number
(default - 1)
Short form: :SENS?
Function: This query reads the sensor which has been selected by the RANGE
command.
Response: A decimal, pressure reading in the current units.
Description
Queries the pressure reading for the selected sensor in the selected units. The sensor can be
changed see, :SENSe[:PRESsure]:RANGe.
Example
TX> :SENS1:PRES?
RX> :SENS:PRES 3616.9282227
TX> :Sens2:Pres?
RX> :SENS2:PRES 3617.1921387
TX> :SENS2:PRES:RANG?
RX> :SENS2:PRES:RANG "3.50barg"
Currently module 2 sensing range is in gauge mode.
TX> :SENS2:PRES:RANG "4.50bara"
TX> :SENS2:PRES:RANG?
RX> :SENS2:PRES:RANG "4.50bara"
It has been changed to pseudo-absolute mode.
TX> :SENS2:PRES?
RX> :SENS2:PRES 4615.0807447
The reading is also changed to pseudo-absolute mode.
Note:
:SOURe[:PRESsure]:RANGe command does not change ‘SENS?’ reading.
It changes the front-panel display.