EasyManuals Logo
Home>GE>Test Equipment>PACE Series

GE PACE Series User Manual

GE PACE Series
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #49 background imageLoading...
Page #49 background image
PACE Series SCPI Manual
K0472 Revision A 4-21
INST:SENS:CALD
Applicability: PACE1000, PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
:INST:SENS[x]:CALD[y]?
where x (sensor number) = 1, 2. . . . . 8 (default -1) y = 1, 2, 3 . . . . .10 (y times last calibration date)
Short form: :INST:SENS:CALD?
Function: Queries sensor calibration dates.
Response: Returns sensor [x] with [y] calibration date:
where x is the sensor number.
x = 1, 2. . . . . 8
where y is the y
th
calibration date for sensor x.
y = 1, 2, 3 . . . . .10 calibration dates stored for each sensor.
Description
This query returns the following:
Example
This shows that the module 2 control sensor was last calibrated on 21
st
November 2009 and
the previous calibration was on 17
th
November 2009.
x Sensor PACE5000, PACE6000 Sensor PACE1000
1 Module 1: control sensor Internal sensor1
2 Module 1: source pressure +ve Internal sensor 2
3 Module 1: source pressure -ve Internal sensor 3
4 Module 1: barometric range (optional) External IDOS-1
5 Module 2: control sensor External IDOS-2
6 Module 2: source pressure +ve External IDOS-3
7 Module 2: source pressure -ve -
8 Module 2: barometric range (optional) -
TX> :INST:SENS5:CALD1?
RX> :INST:SENS5:CALD 2009, 11, 21
TX> :INST:SENS5:CALD2?
RX> :INST:SENS5:CALD2 2009, 11, 17

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PACE Series and is the answer not in the manual?

GE PACE Series Specifications

General IconGeneral
BrandGE
ModelPACE Series
CategoryTest Equipment
LanguageEnglish

Related product manuals