PACE Series SCPI Manual
K0472 Revision A 4-21
INST:SENS:CALD
Applicability: PACE1000, PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
:INST:SENS[x]:CALD[y]?
where x (sensor number) = 1, 2. . . . . 8 (default -1) y = 1, 2, 3 . . . . .10 (y times last calibration date)
Short form: :INST:SENS:CALD?
Function: Queries sensor calibration dates.
Response: Returns sensor [x] with [y] calibration date:
where x is the sensor number.
x = 1, 2. . . . . 8
where y is the y
th
calibration date for sensor x.
y = 1, 2, 3 . . . . .10 calibration dates stored for each sensor.
Description
This query returns the following:
Example
This shows that the module 2 control sensor was last calibrated on 21
st
November 2009 and
the previous calibration was on 17
th
November 2009.
x Sensor PACE5000, PACE6000 Sensor PACE1000
1 Module 1: control sensor Internal sensor1
2 Module 1: source pressure +ve Internal sensor 2
3 Module 1: source pressure -ve Internal sensor 3
4 Module 1: barometric range (optional) External IDOS-1
5 Module 2: control sensor External IDOS-2
6 Module 2: source pressure +ve External IDOS-3
7 Module 2: source pressure -ve -
8 Module 2: barometric range (optional) -
TX> :INST:SENS5:CALD1?
RX> :INST:SENS5:CALD 2009, 11, 21
TX> :INST:SENS5:CALD2?
RX> :INST:SENS5:CALD2 2009, 11, 17