4 Command and Query Summary
K0472 Revision A 4-10
:CAL:PRES:ZERO:TIME:STAT
Applicability: PACE5000, PACE6000
Command Syntax
:CALibration[x]:[PRESsure]:ZERO:TIME:STATe
where: x (module) = 1 or 2 (default - 1)
Parameter: <<boolean>>
Short form: :CAL:ZERO:TIME:STAT1
Function: Sets timed zero on/off.
Default: 0
0 - sets timed zero to OFF
1 - sets timed zero to ON
Query Syntax
:CALibration[x]:[PRESsure]:ZERO:TIME:STATe?
where: x (module) = 1 or 2 (default - 1)
Short form: :CAL:ZERO:TIME:STAT?
Function: Queries the status of the timed zero ON or OFF.
Response:
:CAL:ZERO:TIME:STAT 1
or
:CAL:ZERO:TIME:STAT 0
Description
This command sets the time period for zero on or off. This query gets the status of the timed
zero (on or off).