4 Command and Query Summary
K0472 Revision A 4-12
INPut
The INPut subsystem shows the state of the logical inputs.
:INP:LOG
Applicability: PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Default:
Query Syntax
:INPut[x]:LOGic?
where: x (module) = 1 or 2 (default - 1),
Short form: :INP:LOG?
Function: Asks for state of switch input within the controller module.
Response: first parameter - 0 = OFF, 1 = ON
second parameter - measured pressure at the time of
switching (snapshot in current pressure
units).
Description
This query returns the state of the switch input within the module and the pressure at time of
switching operations.
Example
:INP:LOG?
:INP:LOG 0, 0.8321209
Current logic OFF, pressure was 0.8321209 in current pressure units when the module was
switched to OFF condition.