PACE Series SCPI Manual
K0472 Revision A 4-125
UNIT
The UNIT sub-system configures the instrument’s pressure measurement units.
UNIT:ALT
Applicability: PACE1000, PACE6000
Command Syntax
UNIT:ALTitude
Parameter: <name>
Short form UNIT:ALT
Function: Selects altitude units.
Query Syntax
UNIT:ALTitude?
Short form: UNIT:ALT?
Function: Queries what altitude units are selected
Response: SUFFIX DATA
Description
This command selects the altitude units. The query gets the altitude units are selected.
Note: If the instrument is not in aero mode it returns an error: -221. Settings conflict.