PACE Series SCPI Manual
K0472 Revision A 4-63
:SENS:PRES:CORR:VOL
Applicability: PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Default:
Query Syntax
:SENSe[x][:PRESsure]:CORRection:VOLume?
where x = 1 or 2 is the module number
(default - 1)
Short form: :SENS:PRES:CORR:VOL?
Function: Ask for the estimated volume of the system connected to the instrument.
Response: Decimal number in litres corresponding to volume.
Description
The instrument calculates and reports the volume of the system connected.
Example
TX> :SENS:PRES:CORR:VOL?
RX> :SENS:PRES:CORR:VOL 0.0150000