4 Command and Query Summary
K0472 Revision A 4-122
:SYST:PASS:CEN
Applicability: PACE1000, PACE5000, PACE6000
Command Syntax
:SYSTem:PASSword:CENable:
Parameter: 2317100
Short form: :SYST:PASS:CEN
Function: Enables calibration, default condition - disabled (calibration is not allowed).
Query Syntax
n/a
Short form:
Function:
Response:
Description
This command enables calibration.