PACE Series SCPI Manual
K0472 Revision A 4-13
:INP:LOG:STAT
Applicability: PACE1000, PACE5000, PACE6000
Note: Only applies to the PACE1000 when the VFC option is fitted.
Command Syntax
n/a
Parameter:
Short form:
Function:
Default:
Query Syntax
:INPut[x]:LOGic:STATe?
where: x (module) = 1 or 2 (default - 1),
Short form: :INP:LOG:STAT?
Function: Asks for state of switch input.
Response: 0 = OFF, 1 = ON
Description
This query returns the state of the switch input.