EasyManuals Logo
Home>GE>Test Equipment>PACE6000

GE PACE6000 Technical Manual

GE PACE6000
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #151 background imageLoading...
Page #151 background image
PACE Series SCPI Manual
K0472 Revision A 4-123
:SYST:PASS:CEN:STAT
Applicability: PACE1000, PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
:SYSTem:PASSword:CENable:STATe?
Short form: SYST:PASS:CEN:STAT?
Function: Gets the status of the calibration password.
Response: 0 - disabled
1- enabled
Description
This query asks if the calibration is enabled or disabled.
Example
It can be enabled by sending command and password
TX> :SYST:PASS:CEN 2317100
TX> :SYST:PASS:CEN:STAT?
RX> :SYST:PASS:CEN:STAT 1
When enabled and a 3-point calibration can be carried out.
It can be disabled again by issuing a command and a password, such as:
TX> :SYST:PASS:CDIS 2317100
TX> :SYST:PASS:CEN:STAT?
RX> :SYST:PASS:CEN:STAT 0

Table of Contents

Other manuals for GE PACE6000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PACE6000 and is the answer not in the manual?

GE PACE6000 Specifications

General IconGeneral
BrandGE
ModelPACE6000
CategoryTest Equipment
LanguageEnglish

Related product manuals