4 Command and Query Summary
K0472 Revision A 4-26
:INST:SENS:ZERO
Applicability: PACE1000
Command Syntax
INSTrument:SENSor:ZERO
where: x = 1, 2…8 is the sensor number.
Parameter: <integer> 1, 2…8
Short form: INST:SENS:ZERO
Function: Performs a zero on the selected sensor specified in the sensor index.
Query Syntax
:INSTrument:SENSor[x]:ZERO?
where: x = 1, 2…8 is the sensor number. It defaults to 1.
Short form: :INST:SENS:ZERO?
Function: Queries sensor zero condition.
Response: Gets zero condition [y] of selected sensor [x].
where x is the sensor number.
x = 1, 2. . . . . 8
y =
Description
The command selects a sensor and performs a zero. The query gets the zero condition after
a zero has been performed on the selected sensor.
0 Zero ok
1 Zero in progress
2 Zero timed out
3 Zero outside limits
4 Zero aborted