EasyManuals Logo
Home>GE>Test Equipment>PACE6000

GE PACE6000 Technical Manual

GE PACE6000
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
4 Command and Query Summary
K0472 Revision A 4-26
:INST:SENS:ZERO
Applicability: PACE1000
Command Syntax
INSTrument:SENSor:ZERO
where: x = 1, 2…8 is the sensor number.
Parameter: <integer> 1, 2…8
Short form: INST:SENS:ZERO
Function: Performs a zero on the selected sensor specified in the sensor index.
Query Syntax
:INSTrument:SENSor[x]:ZERO?
where: x = 1, 2…8 is the sensor number. It defaults to 1.
Short form: :INST:SENS:ZERO?
Function: Queries sensor zero condition.
Response: Gets zero condition [y] of selected sensor [x].
where x is the sensor number.
x = 1, 2. . . . . 8
y =
Description
The command selects a sensor and performs a zero. The query gets the zero condition after
a zero has been performed on the selected sensor.
0 Zero ok
1 Zero in progress
2 Zero timed out
3 Zero outside limits
4 Zero aborted

Table of Contents

Other manuals for GE PACE6000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PACE6000 and is the answer not in the manual?

GE PACE6000 Specifications

General IconGeneral
BrandGE
ModelPACE6000
CategoryTest Equipment
LanguageEnglish

Related product manuals