PACE Series SCPI Manual
K0472 Revision A 4-53
:SENS:PRES:INL
Applicability: PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
:SENSe[x][:PRESsure]:INLimits?
where x = 1 or 2 is the module number
(default - 1)
Short form: :SENS:INL?
Function: Query in-limits value
Response: First parameter - current pressure.
Second parameter - in limit:
0 = not in limits
1 = in limits
Description
This command query gets the in-limits value and the in-limits status:
Example
1
TX> :SENS:PRES:INL?
RX> :SENS:PRES:INL 990.0527344, 0
Example
2
TX> :SENS:PRES:INL?
RX> :SENS:PRES:INL 990.0527344, 1