PACE Series SCPI Manual
K0472 Revision A 4-45
:SENS:MACH:INL
Applicability: PACE6000
Command Syntax
:SENSe[x][:MACH]:INLimits <number>
where x = 1 or 2 is the module number
(default - 1)
Parameter: <decimal> MACH in limits value as % full-scale
Short form: :SENS:MACH:INL <number>
Function: The instrument has a MACH in-limits indicator. This can generate a
service request when the MACH value is within limits for a set time period.
Default: 0.01 % full-scale
minimum 0 % full-scale
maximum 100 % full-scale
Query Syntax
:SENSe[x][:MACH]:INLimits?
where x = 1 or 2 is the module number
(default - 1)
Short form: :SENS:MACH:INL?
Function: Query MACH in-limits value
Response: Decimal representing MACH in-limits value as % full-scale.
Description
Sets the in-limits value.
Note: If the instrument is not in aero mode it returns an error: -221. Settings conflict.