EasyManuals Logo
Home>GE>Test Equipment>PACE6000

GE PACE6000 Technical Manual

GE PACE6000
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #69 background imageLoading...
Page #69 background image
PACE Series SCPI Manual
K0472 Revision A 4-41
SENS:AIRF:QFE
Applicability: PACE1000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
SENSe:AIRField:QFE?
Short form: SENS:AIRF:QFE?
Function: This query reads the QFE value when the instrument is in airfield mode.
Response: A decimal reading in selected units of pressure.
Description
Queries the QFE reading, returns the QFE reading in selected units of pressure when the
instrument is in airfield mode.
Note: QFE is the mean sea level pressure corrected for temperature and adjusted for a specific
site (i.e. a datum such as an airfield).

Table of Contents

Other manuals for GE PACE6000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PACE6000 and is the answer not in the manual?

GE PACE6000 Specifications

General IconGeneral
BrandGE
ModelPACE6000
CategoryTest Equipment
LanguageEnglish

Related product manuals