4 Command and Query Summary
K0472 Revision A 4-36
SENSe
The SENSe subsystem selects, configures and queries the measurement functions of the
instrument.
SENS:ALT
Applicability: PACE1000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
SENSe:ALTitude?
Short form: SENS:ALT?
Function: This query reads the aeronautical sensor.
Response: A decimal, altitude reading in the current aeronautical units.
Description
Queries the altitude reading for the aeronautical sensor in the selected aeronautical units.
Note: If the instrument is not in aero mode it returns an error: -221. Settings conflict.