General procedures
Vivid E80/E90/E95 – Service Manual 4-3
GC091052 Rev. 3
Connect AC (mains) Power to the Vivid E80/E90/E95
Connecting AC Power to the Vivid E80/E90/E95 ultrasound unit,
involves preliminary checks of the power cord, voltage level and
compliance with electrical safety requirements.
1. Ensure that the wall outlet is of appropriate type, and that
the Circuit Breaker is turned off.
2. Uncoil the power cable, allowing sufficient slack so that the
unit can be moved slightly.
3. Verify that the power cable is without any visible scratches
or any sign of damage.
4. Verify that the on-site mains voltage is within the limits
indicated on the rating label near the Circuit Breaker on the
rear of the unit.
5. Connect the Power Cable’s female plug to the Power Inlet at
the rear of the unit.
6. Lock the plug in position with the Retaining Clamp (ACC
Clamp).
7. Verify that the Mains Power Circuit Breaker is in OFF
position, if not, switch it OFF.
8. Connect the Power Cable’s other end (male plug) to a
hospital grade mains power outlet with the proper rated
voltage, and the unit is ready for Power ON/Boot Up.

Table of Contents

Other manuals for GE Vivid E95

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E95 and is the answer not in the manual?

GE Vivid E95 Specifications

General IconGeneral
TypeCardiovascular Ultrasound System
Transducers2D, 3D, 4D
Display22-inch high-resolution OLED monitor
PortabilityCart-based
Imaging Modes2D, 3D, 4D, Color Doppler
Advanced FeaturescSound Architecture
ConnectivityDICOM, USB, Ethernet
ApplicationsCardiology, Vascular, Abdominal

Related product manuals