Back End Processor (BEP) parts replacement
Vivid E80/E90/E95 – Service Manual 8-101
GC091052 Rev. 3
15. Restore the Patient Archive images, Report Archive,
User-Defined Configuration (Customer Presets), and
Service settings.
Verification after BEP or HDD replacement
Perform the following steps to verify that the product is
functioning as intended after this replacement:
1. Verify that all screws that you removed earlier have been
installed.
2. If finished, connect cables and probes removed earlier.
3. Power up the Ultrasound system to verify that it operates as
intended.
4. Check/set Date, Time, and Time Zone, and set the Preset
Region accordingly.
5. Reconnect to the network.
6. If the Ultrasound system was connected to InSite ExC,
restore the service settings backed up prior to the software
loading. Reboot and verify remote connectivity.
7. Confirm Patient Archive images, Report Archive, User
Defined Configuration (Customer Presets), and Service
settings are restored.
8. If DVD is installed: Format a DVD, and backup Customer
Presets to confirm proper DVD write functionality.
Hard Disk Drive (HDD) replacement
Preparations
Before you dispose of the hard drive, make sure you remove
ALL PATIENT DATA from the hard drive, given that the hard
drive is still functional. In some countries, you may be required
to delete all software from the disk before returning the hard
drive to the parts warehouse. Follow your local policies.