General Procedures and Functional Checks
4-2 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
General procedures
Power ON/Boot Up
Warnings
NOTE: Do not cycle the Circuit Breaker ON-OFF-ON in less than five
(5) seconds. When turning OFF the Circuit Breaker, the
Ultrasound system should de-energize completely before
turning the circuit breaker ON.
DANGER
ALWAYS CONNECT THE ULTRASOUND SYSTEM TO A
FIXED POWER SOCKET WHICH HAS THE PROTECTIVE
GROUNDING CONNECTOR.
DANGER
NEVER USE A THREE-TO-TWO PRONG ADAPTER; THIS
DEFEATS THE SAFETY GROUND.
DANGER
ENSURE THAT THE POWER CORD AND PLUG ARE
INTACT AND THAT THE POWER PLUG IS THE PROPER
HOSPITAL-GRADE TYPE (WHERE REQUIRED).
CAUTION
Ultrasound system requires all covers.
Operate this Ultrasound system only when all board covers and
frame panels are securely in place. The covers are required for
safe operation, good Ultrasound system performance, and
cooling purposes.
CAUTION
Use only power supply cords, cables, and plugs provided or
designated by GE.

Table of Contents

Other manuals for GE Vivid E95

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E95 and is the answer not in the manual?

GE Vivid E95 Specifications

General IconGeneral
BrandGE
ModelVivid E95
CategoryMedical Equipment
LanguageEnglish

Related product manuals