Replacement Procedures
8-58 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Install the Handle, Left Top / Handle Right Top
Follow this procedure to install the Handle, Left Top / Handle
Right Top:
1. Remove the Gel Cups (if present).
2. Install the Handle, Left Top / Handle, Right Top by pressing
them into the UI Frame, Lower.
3. Install the Gel Cups you removed in step 1.
4. Install the Operating Panel, Lower.
5. Install the Operating Panel Upper.
Palm Rest ASSY replacement
Read and follow
Preparations
When preparing for the replacement, you must perform the
following steps:
1. Power down the Ultrasound system.
2. Disconnect the Mains Power Cable from the wall outlet.
3. Disconnect all probes and external I/O cabling.
4. Remove the Rotary buttons below the Touch screen and the
Volume button.
5. Remove the Operating Panel, Upper.
6. Remove the Operating Panel, Lower. Place it on a clean
surface with the front down.
WARNING
Energy Control and Power Lockout for Vivid E80/E90/E95.
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait for at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
Ultrasound System components may be energized.

Table of Contents

Other manuals for GE Vivid E95

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E95 and is the answer not in the manual?

GE Vivid E95 Specifications

General IconGeneral
BrandGE
ModelVivid E95
CategoryMedical Equipment
LanguageEnglish

Related product manuals