Preparing for setup
Vivid E80/E90/E95 – Service Manual 3-17
GC091052 Rev. 3
Preparing for setup
Verify customer order
Compare items received by the customer to those that are listed
on the delivery order. Report any items that are missing, back
ordered, or damaged.
Physical inspection
Verify that the Vivid E80/E90/E95 arrived intact (visual
inspection).
If the Vivid E80/E90/E95 has been damaged, please refer to
‘Damage in transportation’ on page i-15 in the beginning of this
manual.
EMI protection
The Vivid E80/E90/E95 has been designed to minimize the
effects of Electro-Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the
Vivid E80/E90/E95 from image artifacts caused by this
interference. For this reason, it is imperative that all covers and
hardware are installed and secured before the Vivid E80/E90/
E95 is put into operation.
See ‘EMI limitations’ on page 2-7 for more information about
EMI protection.