EasyManuals Logo
Home>GE>Medical Equipment>Vivid E95

GE Vivid E95 User Manual

GE Vivid E95
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #50 background imageLoading...
Page #50 background image
Introduction
1-20 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Electrostatic discharge (ESD) prevention
WARNING
DO NOT touch any boards with integrated circuits prior to
taking the necessary ESD precautions.
Always connect yourself, via an arm-wrist strap, to the advised
ESD connection point located on the rear of the Ultrasound
system (near the power connector).
Follow general guidelines for handling of electrostatic sensitive
equipment.
WARNING
Risk of electrical shock, Ultrasound system must be turned off.
Avoid all contact with electrical contacts, conductors and
components. Always use non-conductive handles designed for
the removal and replacement of ESD sensitive parts. All parts
that have the potential for storing energy must be discharged or
isolated before making contact.
WARNING
If the covers are removed from an operating Vivid E80/E90/
E95, some metal surfaces may be warm enough to pose a
potential heat hazard if touched, even while in shutdown mode.

Table of Contents

Other manuals for GE Vivid E95

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E95 and is the answer not in the manual?

GE Vivid E95 Specifications

General IconGeneral
BrandGE
ModelVivid E95
CategoryMedical Equipment
LanguageEnglish

Related product manuals