EasyManuals Logo
Home>GE>Medical Equipment>Vivid E95

GE Vivid E95 User Manual

GE Vivid E95
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #254 background imageLoading...
Page #254 background image
Replacement Procedures
8-32 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Disconnect the cables from the Bulkhead board
Follow these steps to disconnect the cables from the Bulkhead
board:
1. Disconnect the Power/USB PLUG from the Bulkhead board.
2. Bend the HDMI Connector Lock upwards, then disconnect
the HDMI plug.
Remove the LCD Arm assembly
Follow these steps to remove the LCD Arm assembly:
1. Slide the LCD Mount Lock Handle into unlocked position.
2. Move the LCD Arm from side to side when at the same time
pulling upwards, until you can lift LCD Arm assembly away.
Install the LCD Arm assembly
Follow these steps to install the LCD Arm assembly:
1. Carefully install the LCD Arm assembly into position, first
feeding the LCD Arm cables down through the Top Console
opening.
2. Push the LCD Mount Lock Handle into locked position.
3. Connect the LCD cables to the connectors on the Bulkhead.
4. Install the Bulkhead Cover.
5. Install the LCD monitor.
Verification
Perform the following steps to verify that the product is
functioning as intended after this replacement:
1. Verify that all screws that you removed earlier have been
installed.
2. If finished, connect the cables and probes you removed
earlier.
3. Power up the Ultrasound system to verify that it operates as
intended.
4. Move the LCD Monitor Arm from side to side and ensure
that it moves as intended.

Table of Contents

Other manuals for GE Vivid E95

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E95 and is the answer not in the manual?

GE Vivid E95 Specifications

General IconGeneral
BrandGE
ModelVivid E95
CategoryMedical Equipment
LanguageEnglish

Related product manuals