Replacement Procedures
8-68 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Remove the XYZ Control Assembly
Follow this procedure to remove the XYZ Control Assembly:
1. Disconnect the cables to the XYZ Control Assembly.
2. Unscrew and remove two (2x) screws.
3. Pull the top of the XYZ Control Assembly forwards, up and
away from the bracket.
Install the XYZ Control Assembly
Follow this procedure to install the XYZ Control Assembly
1. Thread the XYZ Control Assembly into position.
2. Install the three screws at the top, rear. (M6x16 sunken.
Torque: 9.8 Nm.)
3. Install the two remaining screws on the top. (M4x8, sunken.
Torque: 4.9 Nm.)
4. Connect the cables to the XYZ Control Assembly.
5. Install the Top Cover.
6. Install the Rear Cover.
7. Install the Left Side Cover.