To degauss the display
If the mainframe has been subjected to strong magnetic fields, the CRT might
become magnetized and display data might become distorted. To correct this
condition, degauss the CRT with a conventional, external-television-type degaussing
coil.
To clean the logic analysis system
With the instrument turned off and unplugged, use mild soap and water to clean the
front and cabinet of the system. Harsh soap might damage the water-base paint.
To test the logic analysis system
The logic analysis system does not require calibration or adjustment.
If you require a test to initially accept the operation, perform the self-tests in
chapter 3, "Testing Performance."
If the logic analysis system does not operate correctly, go to the beginning of
chapter 5, "Troubleshooting."