Technical Specification
The above specifications may be subject to change without prior notice.
1. Internal absorption current and internal absorption power mean the
absorption current and power of the standard power dissipater inside the power
supply.
2. Resistance programming accuracy differs based on different output current.
For example, 10A Transient 1.0Ω, accuracy: (1.0Ω * 0.25%) +(375mΩ*A /10A)
= 40mΩ.1.0Ω, accuracy: (1.0Ω * 0.25%) +(375mΩ*A /10A) = 40mΩ.
3. Set value accuracy means the set accuracy realized through the panel button
or communication instructions. When external analog is used for programming,
the programming accuracy is 2%FS.
4. Read-back value accuracy means the read-back accuracy realized through
panel display or communication instructions; when external analog is used for
monitoring, the monitoring accuracy is 1%FS.
5. Up time and down time mean the time of establishment from one value to
another value under ON status when the internal standard power dissipater is
enabled.
6. Use 12V/120AH battery for test.
7. When used in parallel, to balance power of each phase, distribute the
instrument to three phases. But must ensure that single machine input voltage
meets Specification requirements.
8. It means the maximum phase current value under minimum work input
voltage.
5.1.6 IT6533C
Parameter IT6533C Ver:V1.6
Output Rating ( 0 ℃-40 ℃)
Output Voltage
0~200V
Output Current
0~120A
Internal Sink
Current
1
0~50A
Output Power
0~6KW
Internal Sink Power
1
0~300W
Output Resistance
Range
0~6.666Ω
Accuracy
2
0.25%+100mΩ
Resolution 1mΩ
Line regulation
±(%of Output+Offset)
Voltage
≤0.01%+20mV
Current ≤0.01%+60mA
Copyright © Itech Electronic Co., Ltd.
106