IEEE-488 Conformance Information
E-3
Table E-2
Coupled commands
Command Also changes To If
:SYST:AMET BURS ... :RANG:AUTO
... :NPLC
... :APER
... :DIG:AUTO
... :DIG
... :AVER:AUTO
... :AVER:TCON
:INIT:CONT
:SYST:AZER:STAT
:ARM:LAY1:COUN
:ARM:LAY2:COUN
:TRIG:COUN
:TRACe:EGR
:TRACe:FEED
:TRACe:FEED:CONT
OFF
0.01
0.01 / line-frequency
OFF
4
OFF
MOV
OFF
OFF
1
1
TRACe:POIN
COMP
CALC1
NEXT
TRACe:FEED is NONE
:SYST:AMET:ASTR
and
:SYST:AMET:SSTR
:SYST:AZER:STAT
:SYST:LSYN:STAT
:CALC:STAT
... :RANG:AUTO
... :NPLC
... :NPLC:AUTO
... :DIG
... :DIG:AUTO
... :AVER:STAT
... :AVER:AUTO
... :REF:STAT
OFF
OFF
OFF
OFF
0.01
OFF
4
OFF
OFF
OFF
OFF
:TRIG:COUN :TRAC:POIN:AUTO
:TRAC:POIN
:TRAC:FEED:PRET:AMO:READ
OFF
TRIG:COUN
TRIG:COUN *
TRAC:FEED:PRET:A
MO:PERC / 100
TRAC:POIN:AUTO is
ON and TRIG:COUN is
an illegal buffer size
TRAC:POIN:AUTO is
ON
TRAC:POIN:AUTO is
ON
:TRAC:FEED:CONT :TRIG:SOUR TIM TRAC:FEED:CONT is
PRET and TRIG:SOUR
is neither IMM nor TIM
:TRAC:POIN :TRAC:POIN:AUTO
:TRAC:FEED:CONT
:TRAC:FEED:PRET:AMO:READ
:TRIG:COUN
OFF
NEV
TRAC:POIN * TRAC
:FEED:PRET:AMO
:PERC / 100
TRAC:POIN SYST:AMET is BURS