1-10 Getting Started
Display modes
For voltage and current readings, there are four display modes described as follows:
• ACTUAL V AND I — This
display mode is used to read back the actual output voltage
and current. This display mode is the RST default. (See Section 2 for details.)
• DVM INPUT — This mode is used to
display the DC voltage applied to the DVM input
of the power supply. (See Section 2 for details.)
• PULSE CURRENT — This
mode is used to display high, low, or average pulse-current
measurements. (See Section 3 for details.)
• LONG INTEGRATION — This mode
is used to display average current measurements
of a pulse or pulses measuring periods between 850msec to 60sec (60 Hz line frequency)
and 840msec to 60sec (50 Hz line frequency). (See Section 4 for details.)
Any one of the four display modes can be the power-on default. Use the SAVE SETUP item
of the
MENU to save the selected display mode in memory, and use the POWER ON SETUP
item to specify the power-on setup (see “Setups — Save, Power-on, and Recall” on page 1-14
for details). A display mode is selected as follows:
1. Press the DISPLAY key and use the
▲ or ▼ key to display the desired mode: ACTUAL
V AND I, DVM INPUT, PULSE CURRENT, or LONG INTEGRATION.
DISPLAY TYPE #1 or DISPLAY TYPE #2 will be shown on the top line of the display.
NOTE DISPLA
Y TYPE #1 is the display mode for the Battery Channel while DISPLAY TYPE
#2 is the display mode for the Charger Channel.
2. Toggle active channel using the
▲
or
▲
keys.
NOTE If activ
e channel is changed back to the original channel, the initial settings are
displayed.