Basic Power Supply Operation 2-13
NOTE Refer to the Programming syntax paragraph of Section 6 for a description of
parameters (e.g., <b>, <NRf>, etc.).
Command notes (outputting voltage and current)
SENSe[1]:CURRent:RANGe <n> Applies to battery channel (#1)
SENSe2:CURRent:RANGe <n> Applies to charger channel (#2)
After specifying a current value, the instrument will go to the most sensitive range to
acc
ommodate that reading. For example, if you are expecting a maximum current
reading of 750mA, you can let <n> = 0.75 (or 750e-3) to select the 5A range. Using the
:RANGe command to manually select a current range disables auto range. Another way
to select a range is to use the MINimum, MAXimum, and DEFault parameters as follows:
SENS:CURR:RANG MIN Select the low current range (5mA) for battery
channel (#1).
SENS2:CURR:RANG MAX Select the high current range (5A) for charger
channel (#2).
SENS2:CURR:RANG DEF Select the default current range for charger channel
(#2).
The response for :RANGe? query returns the selected range value which is either 5.0000
or 0.0050.
SENSe[1]:CURRent:RANGe:AUTO <b> Applies to battery channel (#1)
SENSe2:CURRent:RANGe:AUTO <b> Applies to charger channel (#2)
This command is coupled to the :RANGe <n> command. When auto range is enabled,
the response
for :RANGe? query returns the selected range value which is either 5.0000
or 0.0050. If you then disable auto range, the instrument will remain at the last selected
range.
[SOURce1]:VOLTage <n> Applies to battery channel (#1)
SOURce2:VOLTage <n> Applies to charger channel (#2)
This command sets voltage amplitude in volts: 0 to 15 (1mV resolution).
[SOURce1]:CURRent <n> Applies to battery channel (#1)
SOURce2:CURRent <n> Applies to charger channel (#2)
• With the 5mA measurement range selected, the maximum current limit is 1A.
• Sending a value that exceeds 1A is rejected, and the following message is displayed
briefly:
CURRENT LIMIT ON
mA RANGE ≤ 1A
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2306

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2306 and is the answer not in the manual?

Keithley 2306 Specifications

General IconGeneral
BrandKeithley
Model2306
CategoryPower Supply
LanguageEnglish

Related product manuals