EasyManuals Logo
Home>Keithley>Power Supply>2306

Keithley 2306 User Manual

Keithley 2306
348 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
2-22 Basic Power Supply Operation
DVM measurements
The following command sequence demonstrates how to measure voltage applied to the DVM
input of the power supply:
Battery channel (#1)
DISP:CHAN 1 ‘ Set active channel - battery.
SENS:FUNC ‘DVM’ ‘ Select the DVM Input function.
SENS:NPLC 6 ‘ Set integration rate to 6 PLC.
SENS:AVER 10 ‘ Set average reading count to 10.
READ:ARR? ‘ Trigger and return 10 readings. The average
of the 10 readings is displayed on the front
panel.
Charger channel (#2)
DISP:CHAN 2 ‘ Set active channel - charger.
SENS2:FUNC ‘DVM’ ‘ Select the DVM Input function.
SENS2:NPLC 3 ‘ Set integration rate to 3 PLC.
SENS2:AVER 8 ‘ Set average reading count to 8.
READ2:ARR? ‘ Trigger and return 8 readings. The average
of the 8 readings is displayed on the front
panel.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2306

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2306 and is the answer not in the manual?

Keithley 2306 Specifications

General IconGeneral
BrandKeithley
Model2306
CategoryPower Supply
LanguageEnglish

Related product manuals