SECTION 4
IEEE-488
Reference
4.2.5
F - V/I Ohms
Purpose
To control the V/I ohms function of the Model 487.
Format
Parameters
Fn
FO Disable V/I ohms
Fl Enable V/I ohms
Default
Upon power-up or after receiving a DCL or SDC command, the instrument returns to the op-
tion of the F command that was saved as the default condition.
Description
For V/I ohms, resistance is automatically calculated (and displayed) from the applied voltage
and the measured current in accordance with Ohms Law: R = V/I.
For complete
information
on using V/I ohms, refer to paragraph 3-7.
Programming
OUTPUT 722; “FIX”
Examples
OUTPUT 722; ‘TOX”
! Enable V/I ohms.
! Disable V/I ohms.