This subroutine measures the substrate current when the gate voltage (V
GS
) is swept with V
DS
and
V
BS
held constant. Maximum current measured is returned as the function result. The gate voltage at
maximum current is also returned. In some cases, peak substrate current is referred to as I
SX
.
A substrate connection is mandatory; the ismax parameter returns -1.0 if the sub parameter is less
than 1.
The typical value for the npts parameter is 10 to 20.
V/I polarities
N-channel +V
DS
, +V
low
and V
high
, -V
BS
P-channel -V
DS
, -V
low
and V
high
, +V
BS
Source-measure units (SMUs)
SMU1: Forces vds, maximum current limit
SMU2: Sweeps V
GS
, default current limit
SMU3: Forces vbs, default current limit, measures I
SX
Example
isubmx(d, g, s, sub, vds, vbs, vlow, vhigh, npts, &ismax, &vgmax)
Schematic
kdelay
This subroutine provides an appropriate delay time based on current and voltage values.
Usage
void kdelay(int npin, double i, double v)
npin
Input
i
Input
v
Input